Delivery checklist
1 Connecting cable
1 Hexagonal pin-type spanner
1 Operating instructions
1 Notes on moving the equipment safely
The rotor(s) and associated accessories are included in the delivery in the quantity ordered.
8
Unpacking the centrifuge
Lift the carton upward and remove the padding.
Do not lift by the front panel.
Observe the weight of the centrifuge, refer to chapter "Technical specifications".
Lift the centrifuge on both sides with an appropriate number of helpers and place it on the laboratory table.
9
Initial operation
Remove the transportation safety device from the bottom of the housing, see sheet "Transportation safety device"
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specification".
Turn on the mains switch. Switch position "".
The machine type and program version will be displayed and the LEDs light up. After 8 seconds,
OPEN
OEFFNEN is displayed and the left LED on the
Open the lid.
The last used centrifuge data will be displayed.
36/99
key blinks.
STOP / OPEN
Rev. 10 / 04.2020