9.3 Cord test

The cord test can be used both for the testing of IEC power cords (device connecting cables
with IEC coupler) and for the testing of cable reels, multiple distributors and extension cables.
9.3.1

Testing of IEC power cords (IEC adapter cables)

-
Disconnect the plug of the mains connection cable from the socket K of the BENNING ST 725.
-
Connect the IEC power cord to be tested to the BENNING ST 725 by means of the IEC
connector J.
-
Press the
-key 2 to start the automatic testing procedure.
-
The test starts with measuring the protective conductor resistance R
-
Depending on whether the value is higher or lower than the limiting value, a
indicated next to the R
The protective conductor resistance depends on the length and cross-section
of the line to be tested.
-
If R
is higher than the admissible limiting value (≤ 0.3 Ω up to a length of 5 m) but lower
PE
than 1 Ω, the measured value will be displayed without a rating, the "tAble" symbol will
appear on the display and the test procedure will be stopped. The responsible testing staff
determines by means of the limiting value table (see section 7.10 or the table on the rear
of the BENNING ST 725) and by means of the line length of the test object whether the
displayed measured value is acceptable or not.
Press the
-key 2 to give the measured value a positive rating and a
next to the R
PE
Press the
-key 3 to give the measured value a negative rating and a
next to the R
PE
been stopped.
-
Please refer to Table 1 for typical resistance values of lines.
Length
5 m
10 m
25 m
50 m
Table 1:
Resistance values of the protective conductor depending on length and cross-section
-
After the R
test has been passed, the measurement of the insulating resistance is carried
PE
out automatically.
-
Depending on whether the value is higher or lower than the limiting value, a
indicated next to the R
-
After the R
test has been passed, the external conductor (L) and the neutral conductor
ISO
(N) are checked for line breaks and short-circuits. A passed test regarding line breaks and
short-circuits is indicated by a
-
The "PASS" symbol confirms successful testing of the entire testing procedure.
-
If the test regarding line breaks and short-circuits has failed, one of the following symbols is
indicated instead of the "Good" symbol:
-
"OPEN" symbol:
07/ 2018
-symbol.
PE
symbol. The test procedure will be continued.
symbol. "FAIL" appears on the display to confirm that the measurement has
1.0 mm²
0.1 Ω
0.2 Ω
0.5 Ω
1.0 Ω
symbol.
ISO
BENNING ST 725
Cross-section
1.5 mm²
0.06 Ω
0.12 Ω
0.3 Ω
0.6 Ω
next to the
and the "Good" symbol.
.
PE
or a
will be displayed
will be displayed
2.5 mm²
0.04 Ω
0.08 Ω
0.2 Ω
0.4 Ω
or a
is
is