English
Please read these instructions carefully to
familiarise yourself with the operation.

Product description

The SR Ivocap System is an injection technique, which has been
especially developed to compensate curing shrinkage.
Heat/pressure curing allows shrinkage of the acrylate during
polymerization to be compensated by inflowing material.
The SR Ivocap injection technique, therefore, permits the
fabrication of high quality products with excellent physical
properties.
Composition
SR Ivocap High Impact
Powder:
Polymethyl methacrylate, copolymer, benzoyl peroxide,
pigments
Liquid:
Methyl methacrylate, dimethacrylate (cross-linking agent),
copolymer
SR Ivocap Clear
Powder:
Polymethyl methacrylate, dibenzoyl peroxide
Liquid:
Methyl methacrylate, dimethacrylate (cross-linking agent) ,
stabilizer
List of parts
SR Ivocap System
A) Pressure apparatus "ID2"
A1 = Manometer
A2 = Extrusion piston
A3 = Vise shank
A4 = Safety loop
B) Clamping frame
B1 = Ratchet
B2 = Pressure ring with lever
B3 = Pressing plate
C) Flask
C1 = Flask half M
C2 = Flask half W
C3 = Flask cover
C4 = Injection funnel
D) Capsule plunger
E) Investment aid
F) Funnel
G) Timer
2
Indication
SR Ivocap High Impact, SR Ivocap Clear
Complete dentures
Partial dentures
Bases and relines
Orthodontic appliances
Bite guard splints
Contraindication
Direct intraoral contact of unpolymerized material.
If the patient is known to be allergic to any of the
components listed.
Important notes
Do not wet teeth with SR Ivocap monomer, as this will
weaken the bond to the tooth
Soak models thoroughly before the investment
Thoroughly isolate inner flask surfaces before the
investment
Injection channels must have a diameter of 3–5 mm
Thoroughly clean and isolate plaster surfaces
Observe polymerization and cooling times precisely
Contact of solvents or monomer with the polymerized
denture base material can cause white discolouration
If the cervical area is to be isolated, Ivoclar Vivadent
recommends using A silicones with a Shore A hardness
of >65 (e.g. Flexistone Plus/Virtual
Condensation-reaction silicones (C silicones) may have
a negative effect on the denture base material. Other
products may cause negative reactions of the denture
base material.
The monomer composition of SR Ivocap HI, SR Ivocap
Clear and SR Ivocap Elastomer is not identical. Avoid a
mix up of these three monomers as it causes product
defects.
Warnings
The monomer contains methyl methacrylate (MMA)
MMA is highly flammable and irritating
(flash point: +10 °C / 50 °F).
MMA irritates the eyes, respiratory organs and skin
Contact may evoke sensitive skin reaction.
Avoid contact of skin with monomer and unpolymerized
material. Commercial medical gloves do not provide
protection against the sensitizing effect of methacrylates.
Do not inhale vapours
Keep the material away from sources of heat – do not
smoke
Do not dispose of the monomer in the drainage system
Avoid electrostatic charging
Heavy Body Fast).
®