15.3 Use of the microscope (B-510LD series)
The filter turret is provided with 4 positions.
•
In each of the four positions a fluorescence filter can be in-
serted, which can be selected from the options shown in the
table below.
•
You can always add an additional filter after the first in-
stallation (see section 8.6).
•
If all four turret positions are full, the observation in
transmitted light will be affected by the presence of the
fluorescence filter.
1. Turn on the main switch ①. (Fig. 100)
•
For the model B-510LD4D: turn on the microscope using
the main switch on the right side of the frame. (Fig. 53)
2. Move the filter selector ② to one of the available positions
until the click stop. (Fig. 101)
•
When the filter is in the correct position, the dedicated LED
lights up.
3. Adjust to the desired brightness by rotating the dial ①.
•
When switching the fluorescence filter, the LED light
goes out. This is not a defect.
FILTER
EXCITATION
NAME
FILTER
M-1223
325-375 nm
M-1223.1
340-390 nm
M-1222
390-420 nm
M-1220
455-495 nm
M-1220.1
455-495 nm
M-1221
510-550 nm
M-1221.1
510-550 nm
M-1228
582-603 nm
M-1224 (*)
590-650 nm
M-1225 (*)
595-645 nm
M-1226 (*)
623-678 nm
M-1227 (*)
720-760 nm
(*) If the use of a camera is needed, please order it by specifying with "AR GLASS" in order to observe above
650nm.
DICHROIC
EMISSION
MIRROR
FILTER
415 nm
435LP nm
405 nm
420-470 nm
440 nm
450LP nm
500 nm
510LP nm
500 nm
518-542 nm
570 nm
575LP nm
570 nm
585-625 nm
610 nm
615-645 nm
660 nm
665LP nm
655 nm
665-715 nm
685 nm
690-750 nm
770 nm
780LP nm
Page 53
②
APPLICATIONS
•
DAPI, Hoechst, NucBlue, Alexa Fluor 350, BFP, Cou-
marin
•
Pacific Blue, Spectrum Blue
•
GFP, Alexa Fluor 488, Calcein, SYBR Green, FITC,
Fluo-4, MitoTracker Green
•
Spectrum Gold, Propidium Iodide, Ethidium Homodi-
mer I
•
Texas Red, Alexa Fluor 594, mRaspberry, Zombie Red
•
Cy5, Alexa Fluor 647, DRAQ5
•
Alexa Fluor 660, DRAQ5
•
Indotricarbocyanine, DiR
①
F ig. 100
F
ig. 100
F ig. 101
F
ig. 101