6.3.5 Check the Cooling Unit parts, the Filling Chamber Screen and
the Mouthpiece
X Check all parts for damaged surfaces, cracks, softening, hardening,
embrittlement, contamination or discoloration.
X Replace damaged parts.
6.3.6 Assembling the Cooling Unit
CAUTION
Plant particles in the Cooling Unit
Clogging of the Cooling Unit
Inhalation of plant particles
• Insert the Screen correctly into the Cooling Unit Cap.
X Assemble parts in reverse order as described in chapter 6.3.1 Disassemb-
ling the Cooling Unit with Mouthpiece.
X Insert the small Base Seal Ring completely into the Cooling Unit Cap.
Ö The small Base Seal Ring is no longer visible.
X When assembling, make sure that Seal Rings are properly seated.
X Insert Screen in the groove provided for this purpose.
6.4 Filling Chamber
6.4.1 Cleaning the Filling Chamber
NOTICE
Material damage caused by improper cleaning
Penetration of isopropyl alcohol into the Vaporizer
• Do not moisten cotton swabs too much.
X Clean the Filling Chamber with a cotton swab soaked in isopropyl alcohol.
6.4.2 Removing the Screen
X If a Screen is clogged, it must be replaced.
X Press in the Screen, using the Filling Chamber Tool along the groove of
the Filling Chamber.
Ö The Screen bulges and detaches from the groove.