INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MIGHTY + MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY + MEDIC Vaporizer should assure that it is used in such an environ-
ment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
© Copyright Gebrauchs.info
IEC 60601
test level
3 Vrms
150 kHz to 80
MHz
6 Vrms
ISM/Amateur
Radio bands
inside 150 kHz
to 80 MHz
3 V/m
80 kHz to 2.7
GHz
RF
communication
equipment
inside 80 MHz
to 6 GHz
A-5
Compliance
level
Portable and mobile
RF communications
equipment should
be used no closer to
any part of the
MIGHTY + MEDIC
Vaporizer, including
cables, than the
recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Recommended
3 Vrms
separation
150 kHz to 80
distance:
MHz
6 Vrms ISM
d = 1.2 √P
Radio bands
inside 150 kHz
to 80 MHz
3 V/m
d = 1.2 √P
80 kHz to 2.7
80 MHz to 800 MHz
GHz
RF
d = 2.3 √P
800 MHz to 2.7 GHz
communication
equipment
inside 80 MHz
to 6 GHz
Electromagnetic
environment –
guidance