Télécharger Imprimer la page

Vapormed Mighty+ MEDIC Mode D'emploi page 251

Publicité

INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
2 Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The MIGHTY + MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY + MEDIC Vaporizer should assure that it is used in such an environ-
ment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
© Copyright Gebrauchs.info
IEC 60601
test level
± 8 kV contact
discharge
± 15 kV air
discharge
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Compliance
level
± 8 kV contact
Floors should be
discharge
wood, concrete or
ceramic tile. If
± 15 kV air
floors are covered
discharge
with synthetic
material, the
relative humidity
should be at least
30%.
± 2 kV for
Mains power quality
power supply
should be that of a
lines
typical residential,
commercial or
± 1 kV for
hospital
input/output
environment.
lines is not
applicable
Mains power quality
± 1 kV line(s) to
should be that of a
line(s)
typical residential,
± 2 kV line(s) to
commercial or
earth
hospital
environment.
A-3
Electromagnetic
environment –
guidance

Publicité

loading

Produits Connexes pour Vapormed Mighty+ MEDIC