5 C O M M I S S I O N I N G A N D O P E R AT I O N
5.9.4 Base Temperature
X Switch on the Vaporizer.
X Press the plus or minus button.
Ö Permanent change of temperature setting between 40 °C and 210 °C
(104 °F and 410 °F).
X Press the plus button briefly.
Ö Increasing the setpoint value in 1 °C (1 °F) steps.
X Press the minus button briefly.
Ö Reducing the setpoint value in 1 °C (1 °F) steps.
X Press the plus or minus button permanently.
Ö Continuous increase or decrease of the target value.
Ö Display of the current temperature (actual temperature) on the display in
the upper line.
5.9.5 Booster Temperature
X Switch on the Vaporizer.
X Double-click on the orange ON/OFF switch.
X Press the plus or minus button.
Ö Permanent adjustment of the booster temperature offset (value range 1 °C
to 170 °C (1 °F to 306 °F)).
Ö A double vibration and a slowly flashing setpoint temperature display
signal that the booster temperature has been reached.
Ö Booster mode ends automatically 90 seconds after the booster tempera-
ture has been reached; the set temperature is then the basic temperature
again.
5.9.6 Superbooster Temperature
X Switch on the Vaporizer.
X Click the orange ON/OFF switch three times.
X Press the plus or minus button.
Ö Permanent adjustment of the booster temperature offset (value range 1 °C
to 170 °C (1 °F to 306 °F)), as the superbooster temperature offset is fixed.
Ö A double vibration and a quickly flashing setpoint temperature display
signal that the superbooster temperature has been reached.
Ö Superbooster mode ends automatically 90 seconds after the booster
temperature has been reached; the set temperature is then the basic
temperature again.
© Copyright Gebrauchs.info