INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Recommended Separation Distances between portable and mobile RF
Communications Equipment and the MIGHTY + MEDIC Vaporizer
The MIGHTY + MEDIC Vaporizer is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the MIGHTY + MEDIC Vaporizer can help prevent elec-
tromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and
the MIGHTY + MEDIC Vaporizer as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the rec-
ommended separation distance d in meters [m] can be determined using the
equation applicable to the frequency of the transmitter, where P is the max-
imum output power rating of the transmitter in watts [W] according to the
transmitter manufacturer.
Note 1:
Note 2:
Medical electrical equipment is subject to special precautions with respect to
electromagnetic compatibility.
The relevant verification according to EN 60601-1-2 is available.
© Copyright Gebrauchs.info
Separation distance according to frequency of
150 kHz –
80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagne-
tic propagation is affected by absorption and reflection from
structures, objects and people.
transmitter [m]
80 MHz –
800 MHz – 2.5 GHz
800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
A-7
d = 2.3 √P
0.23
0.73
2.3
7.3
23