6.3.2 Remove deposits in Cooling Unit and Mouthpiece
X Remove coarse deposits of plant particles inside the Cooling Unit with the
Filling Chamber Tool.
CAUTION
Isopropyl alcohol 70% (V/V) is flammable and may easily ignite
Danger of burns
• Comply with the manufacturer's safety instructions for isopropyl
alcohol 70% (V/V).
NOTICE
Material damage caused by improper cleaning
Bleaching of the Cap Lock Sticker.
Dissolving the adhesive of the Cap Lock Sticker.
• Wipe off the Cap Lock with Signet hologram only with isopropyl
alcohol 70 % (V/V).
NOTICE
Material damage caused by improper cleaning
Embrittlement of plastic parts.
• Do not immerse plastic parts in isopropyl alcohol 70% (V/V) for more
than 30 minutes.