P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the mattress before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
Brake mechanism works
Steer function works
Siderails raise, lower, and latch
Casters lock when you apply the brakes
Casters lock and swivel
Casters are free of wax and debris
Fowler backrest raises, lowers, and latches
Gatch raises, lowers, and latches
Skins are not cracked
Trendelenburg/Reverse Trendelenburg raises and lowers from all locations
IV pole option is intact and latches in all positions
Oxygen bottle holder option is intact and opens and closes
No damage to the head and foot trays
Ground chain intact
Body restraints option latch and are secure
No rips or cracks in the mattress cover
Accessories and mounting hardware in good condition and work
No leaks at hydraulic connections
Hydraulic jacks hold
Hydraulic drop rate set
Hydraulic oil level sufficient
Lubricate where required ( Lubrication points (page 28))
Product serial number:
Completed by:
Date:
L L u u b b r r i i c c a a t t i i o o n n p p o o i i n n t t s s
With the Fowler backrest at 0 degrees, apply Syntech grease (3000-200-719) through the slot and hole in the crank screw
assembly (Figure 14). Wipe off excess grease.
EN
28
0747-109-005 Rev AC.0