Fault
Er:[No.]:11
is displayed.
Er:[No.]:12
is displayed.
Er:[No.]:16
is displayed.
Er:[No.]:32
is displayed.
10
SERVICING
11
TECHNICAL DATA
11.1
General technical data
Dimensions
• Depth
• Width
• Height (measuring rod not extended)
Weight of measuring station
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
Ambient conditions, transport
• Temperature
• Air pressure
• Humidity
Cause
The device has too high a load or too
high a load in one corner.
The device has been switched on with
too high a load.
The device was caused to oscillate, the
zero point could not be determined.
An internal communication error oc-
curred.
The product needs to be set up carefully and serviced regularly. Depending on
how frequently the product is used, seca recommends servicing at intervals of 3
to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
► You can find your local service partner at www.seca.com.
Remedy
• Remove the load from the device or dis-
tribute the weight more evenly.
• Switch off the device, then switch it back
on again.
• If the problem persists, inform seca Ser-
vice.
• Remove the load from the device.
• Switch off the device, then switch it back
on again.
• Switch off the device, then switch it back
on again.
• Switch off the device, then switch it back
on again.
• If the problem persists, inform seca Ser-
vice.
+10 °C to +40 °C / +50 °F to 104 °F
30 % – 80 %, no condensation
-10 °C to +65 °C / +14 °F to 149 °F
0 % – 95 %, no condensation
-10 °C to +65 °C / +14 °F to 149 °F
0 % – 95 %, no condensation
English
576 mm
410 mm
1356 mm
Approx. 13 kg
700 – 1060 hPa
700 – 1060 hPa
700 – 1060 hPa
Servicing • 31