Technical Data; Data - ProMinent DULCOTEST ICT 2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Technical data

14
Technical data
14.1

Data

General information
Measured variable
Electrolytic conductivity
Applications
Production processes in the chemical industry,
phase separation of product-compounds,
determining concentrations of aggressive chemicals
Technical design
Dimensions (L x W)
247 x 47 mm
Installation depth
148 mm
Weight
approx. 1 kg
Materials
Sensor shaft
PFA, fully extrusion-coated
Seal
Chemraz
®
Sensor characteristic values
Conductivity measurement
range
20 μS/cm ... 2000 mS/cm (compensated)
Measurement accuracy
± (5 μS/cm + 0.5 % of the measured value) at T < 100 °C
± (10 μS/cm + 0.5 % of the measured value) at T > 100 °C
Reproducibility
≤ 0.5 % of the measured value
-1
Cell constant
1.98 cm
± 5 %
Temperature gauge
PT 100, class A, fully extrusion-coated
Temperature compensation
0...100 °C
Electrical connection data
Electrical connection
5 m fixed cable in all versions
Process conditions
Operating temperature range
-10...+125 °C
Operating pressure range
0...16 bar
ProMinent
®
45