15 Technical data
Measured
variable
Hypochlorous acid (HOCl)
Probe may not be used in connection with organic chlorine preparations (e.g.
trichlorocyanuric acid) or stabilisers (e.g. isocyanuric acid)!
Application
range
Swimming pool water, drinking water or water of a similar quality
Measurement
range
(30 ° C, pH 7.2)
CLE 3-mA-0,5 ppm:
CLE 3-mA-2 ppm:
CLE 3-mA-10 ppm:
CLE 3-mA-20 ppm:
CLE 3-mA-50 ppm:
CLE 3-mA-100 ppm:
CLE 3-DMT-5 ppm:
CLE 3-DMT-50 ppm:
CLE 2.2-4P:
pH range
pH 5.5...8.0
Temperature
range
5...45 °C (Temperature compensated)
No temperature leaps!
Max. temperature fluctuation speed < 0.3 °C/min
Storage
temperature
5...50 °C
Conductivity
water sample
50 µS/cm - 10,000 µS/cm
Resolution
Corresponds to lower measurement range threshold
Max. pressure
DGM:
DLG:
Flow
DLG III in-line probe housing A/B and/or DGM
Crosssensitivity
Di-/Trichloramine, iodine, ClO
Operating life
Membrane cap
Typically 1 year depending upon the water quality.
The presence of surfactants will considerably reduce the operating life of the
equipment.
®
ProMinent
0.01...0.5 mg/l
0.02...2 mg/l
0.1...10 mg/l
0.1...20 mg/l
0.5...50 mg/l
1...100 mg/l
0.01...5 mg/l
0.1...50 mg/l
0.1...20 mg/l
1 bar (atmospheric pressure at outlet)
1 bar (atmospheric pressure at outlet)
, ozone and bromine, bromamine
2
Technical data
(nominal slope:
24 mA/ppm)
(nominal slope:
6 mA/ppm)
(nominal slope: 1.2 mA/ppm)
(nominal slope: 0.6 mA/ppm)
(nominal slope: 0.24 mA/ppm)
(nominal slope: 0.12 mA/ppm)
(nominal slope:
65 mV/ppm)
(nominal slope: 6.5 mV/ppm)
(nominal slope: 100 mV/ppm)
No negative pressure!
No negative pressure!
Recommended: 40 - 60 l/h
Minimum:
Maximum:
20 l/h
100 l/h
33