Repair / Disposal / Technical Data
Measured variables H
Applications ranges Oxidation of drinking water, landfill leachate, contaminated ground
Measuring range PER 1-mA-50 ppm:
Resolution In accordance with the lower limit of the instrument measuring
Nominal slope at pH 7, T = 30 °C:
Reaction time t
pH range 1 - 11
42
12

Repair

The sensor can only be repaired in the factory. You should therefore
return it in its original packaging. Prepare the sensor for this
beforehand (as described in section 11 Decommissioning).
13

Disposal

IMPORTANT
Please observe the applicable national regulations.
You may return the decontaminated used equipment to ProMinent
Dosiertechnik GmbH, Heidelberg, against prepaid postage.
14

Technical data

O
(hydrogen peroxide)
2
2
water
Disinfection of cooling, service, and production water in the
pharmaceutical, food and beverages industry as well as in
swimming pools
Deodorization (gas scrubber) in municipal and industrial purification
plants
Dechlorination in chemical processes
Chemical bleaching in the timber, paper, textile and mineral
materials industry
Organic synthesis in the chemical, pharmaceutical, and cosmetics
industry
PER 1-mA-200 ppm:
PER 1-mA-2000 ppm: 20... 2000 mg/l
range
PER 1-mA-50 ppm:
PER 1-mA-200 ppm:
PER 1-mA-2000 ppm:
approx. 8 min
90
0.5...50 mg/l
2... 200 mg/l
240 µA/ppm
60 µA/ppm
6 µA/ppm
ProMinent
®