Installation, electrical

34
6
Installation, electrical
IMPORTANT
• Connect the sensor only when the mains power supply is
switched off.
• Do not shorten the connection lead. The measurement result
will otherwise be incorrect. The warranty is void in this case.
• If using a DULCOMETER
tivity, Identcode characteristic "Electrical connection": 7) use
a mains cable with earth lead! The sensor cable screening
must be earthed by the earth lead if there is potential inter-
ference!
• Install according to the relevant national directives.
Break open the large opening in the middle of the base of
the DULCOMETER
DULCOMETER
®
D1C, part 1)
Screw in a 13.5 skintop connector (supplied with the
®
DULCOMETER
D1C controller)
Cut off the brown and the yellow cable (see Fig. 4)
Earth the total cable screen via the power lead of the mains cable
of the DULCOMETER
ference. For this purpose crimp e.g. a screen connector onto the
total screen.
Connect the sensor to the DULCOMETER
accordance with the terminal connection diagram
Tighten the skintop connector until watertight, ensuring that there
is still sufficient cable inside the DULCOMETER
®
D1C controller inductive conduc-
®
D1C housing (see Operating instructions
®
D1C controller if there is potential inter-
®
D1C controller in
®
D1C controller.
ProMinent
®