Transport and storage
4.1
Transport and storage conditions
Transport conditions
Storage conditions
AB1801en
original RMA form from the manufacturer. The Return Material Authorisation
(RMA) form contains a Declaration of No Objection (UBE), which must be
completed in full and enclosed with the return.
If the device and/or accessories are returned to the manufacturer, the
complete return shipment must be cleaned and decontaminated by the
sender. If returns are not cleaned and/or decontaminated or are insuffi-
ciently cleaned and/or decontaminated, this will be performed by the manu-
facturer and charged to the sender.
The original transport locks must be attached for return shipment, see
Chapter 4 'Transport and storage' on page 15 . The device must be
⮫ 
shipped in its original packaging.
NOTICE
Damage to the device due to failure to use the transport
locks.
Secure the transport locks before transporting the
device.
NOTICE
Damage to the device due to condensation.
There is a risk of condensation forming on electrical com-
ponents when component surfaces are cold and the sur-
rounding air is warmer. The condensation that forms may
cause a short circuit and/or destroy electronics.
Warm the device up for at least 3 hours in a warm room
before connecting it to the mains.
or
Warm up for 30 minutes in a cold room.
Before transporting, fasten the transport lock and disconnect the device
from the mains socket.
The transport temperature must be between -20 °C and +60 °C.
Humidity must not be condensing. Humidity must be between 10% and
80%. 
Be aware of the weight of the device.
When transporting using a transport aid (e.g., a pallet truck), the trans-
port aid must be able to carry at least 1.6 times the transport weight of
the device.
Secure the device to prevent it tipping over and falling down during
transport.
Never transport the device sideways or upside down.
The device must be stored in the original packaging.
Only store the device in dry rooms.
The storage temperature must be between -20 °C and +60 °C.
Humidity must not be condensing. Humidity must be between 10% and
80%. 