Software operation
RCF indicator
24 / 39
[t] buttons to set the desired value.
5.
Use the
Adjustable from 1 to 59 seconds, in 1 second increments.
The parameters t/min and t/sec must be set to zero to set continuous
operation.
'--:--' is displayed.
��
[SELECT] button.
6.
Press the
�� Speed
'RPM' is displayed.
[t] buttons to set the desired value.
7.
Use the
A numerical value from 200 RPM to the maximum rotor speed can be
set.
Adjustable in 10 second increments.
Over 10000 RPM adjustable in increments of 100.
[SELECT] button.
8.
Press the
�� The DEC brake level is displayed.
fast: short ramp-down time
slow: long ramp-down time
[t] buttons to set the desired value.
9.
Use the
[START/PULSE] button.
10.
Press the
�� The settings are stored.
1.
If required: Press the
[RCF] button to toggle between the two parameters
�� Press the
'RPM' ) and RCF ( '>RCF<' ).
RPM (
2.
Press the
[SELECT] button.
't/min' is displayed
�� Runtime in
[t] buttons to set the desired value.
3.
Use the
Adjustable from 1 to 99 minutes in 1 minute increments.
The parameters t/min and t/sec must be set to zero to set continuous
operation.
'--:--' is displayed.
��
[SELECT] button.
4.
Press the
�� Runtime in
't/sec' is displayed.
[t] buttons to set the desired value.
5.
Use the
Adjustable from 1 to 59 seconds, in 1 second increments.
The parameters t/min and t/sec must be set to zero to set continuous
operation.
'--:--' is displayed.
��
[SELECT] button.
6.
Press the
�� Centrifuging radius
[t] buttons to set the desired value.
7.
Use the
A numerical value from 10 mm to 250 mm can be set.
Adjustable in 1 millimetre increments
[SELECT] button.
8.
Press the
�� Relative centrifugal force
Rev.: 10 / 11.2023
[RCF] button to select the RCF indicator.
'RAD/mm' is displayed.
'RCF' is displayed.
AB1801en