INSTRUCTIONS FOR USE
c. Insert the cross pin into the H-bracket, through the
bowtie, and through the other end of the H-bracket.
WARNING:
Do not use the H-bracket without the cross pin
installed correctly. To do so could cause patient fall,
injury, or equipment damage.
d. Make sure the drop lock on the end of the cross pin moves to the side to hold the
H-bracket in position.
e. Repeat steps Step a through Step d on the
opposite bowtie.
f.
Determine which hole location on the H-
bracket to install the support top.
NOTE: The hole locations are numbered on the side of the H-brackets.
WARNING:
Failure to use correct lifting methods could cause patient fall, injury, or equipment
damage.
g. With the help of another person, align the ends of the support top with the pre-
determined hole location.
WARNING:
Never remove a cross pin from a support top while a patient is on the support top.
Patient fall or injury could occur.
h. At each end of the support top, insert a cross
pin into the H-bracket, through the top
connector assembly, and through the other
end of the H-bracket.
Document Number: 80028193
Version: G
Page 44
English
Issue Date: 03 FEB 2023
Ref Blank Template: 80025117 Ver. F