Product overview
The different versions of the test kits are listed below:
Systems for use at high pressure (max. 300 bar)
Aerotest HP
Aerotest HP USA
Aerotest HP NF
Aerotest HP NOx
Aerotest Navy
Systems for use at low pressure (max. 20 bar)
Aerotest Alpha
MultiTest med. Int.
The MultiTest med. Int. is used to measure contamination in
compressed air, laughing gas, carbon dioxide and oxygen, in
accordance with USP (United States Pharmacopeia) and
European Pharmacopeia requirements.
WARNING
The product must (with the exception of MultiTest med Int.) be
used only for breathing air in accordance with legal
regulations and standards such as DIN EN 12021.
28
WARNING
Danger of an explosion or a severe fire if oxygen escapes
Oxygen is highly oxidizing.
Only use the product outside buildings or in very well ventilated
spaces with a maximum O
2
Smoking and naked lights are forbidden in the vicinity of the
product. If an escape of oxygen is suspected, turn off the
product immediately and ensure that the O
the ambient air is no more than 23 %Vol. and that any clothing
worn is well aired for at least 15 minutes. Comply with the
national regulations on handling oxygen.
WARNING
The measured gas disperses freely during the measurement.
The device may only be used in well aired rooms or in the open
air.
Observe the national laws and guidelines for maximum
workplace concentrations. When measuring oxygen, it must
be noted that the maximum concentration of oxygen in the
room must not exceed a value of 23 %Vol.
concentration of 23 %Vol.
concentration in
2