Maintenance
6.1
Maintenance table
Work to be carried
Components
out
Device and
Visual check for
components
damage
Pressure regulator
Check and replace
O-ring
if necessary
Replace
Measuring device
Check flow using
bubble test
Check volume flow
Sintered filter
Replace
First-stage regulator
Major overhaul
of the high-pressure
systems and
measuring device
1)
Visual inspection for correct seating, damage, or contamination
2)
by Dräger Service
6.2
NOTICE
Do not clean the metering valve with soapy water. After
cleaning the measuring device, ensure that each metering
valve has been screwed back into the appropriate connection
on the measuring device.
NOTICE
Oil-free, grease-free and dry air in accordance with
DIN EN 12021 must be used for all cleaning, rinsing and
X
drying operations listed below.
1)
X
1. After each use, blow out the tube holder (rubber part) with
air.
If the measurement indicates that the measured air is
X
contaminated:
X
2. Purge the device with air for at least 6 minutes.
3. In case of severe contamination, remove the measuring
2)
X
device. Clean the individual components (excluding the
metering valves) with soapy water.
X
4. Dry with air.
2)
X
5. Purge the metering valve
with air only in the
direction of the arrow.
6. Mount the measuring
device.
Cleaning
en
Maintenance
39