RS 232-422 Communication Descriptions
Communication format:
When a baud rate of 2400, 4800, 9600 baud is used, the requested information
NOTE
from Host to controller must be sent with a minimum interval of 1 second to
avoid the controller hang-up.
Communication protocol:
Host = Master
Controller = Slave
The communication is performed in the following way:
Figure 16
Turbo-V 1000HT Controller User Manual / 87-900-894-01
8 data bit
no parity
1 stop bit
The baud rate is programmable via front panel from
600 to 9600 baud. The controller is factory-set for 9600 baud
operation.
Technical Information
Optional Serial Port
(G)
17
255/310