Technical Information
Use
and the controller is ready to restart (see paragraph "Startup").
When PUMP LIFE is reset to 000, the CYCLE number is also reset to 000.
NOTE
Speed Adjustment
By pressing CYCLE NUMBER and PUMP CURRENT push button
together for at least five seconds the processor enters in a routine
where it is possible to adjust the rotational speed.
The display shows:
S
Press the PUMP push button to select the digit to be changed and
when press the CYCLE push button to change the value. The speed
value can be changed from 24 to 38 KRPM. Different values are
rejected. The speed adjustment sets automatically the LOW SPEED
frequency to 2/3.
The speed can be adjusted only when the pump is stopped.
NOTE
294/310
P
E
E
D
3
8
Turbo-V 1000HT Controller User Manual / 87-900-894-01