Technical Information
Interconnections
Controller to Pump Connection
A five-meter long cable is to connect the controller to the pump. The
following figures show the controller output connector configuration
where pins:
Figure 13
250/310
A-F = pump temperature sensor
B-C-D = 54 Vac 3-phase output to pump motor stator
E = ground
Controller-to-pump connector (applicable to model 969-9445 and
969-9545)