17
Technical Information
Use
Front / Remote Selection
Press CYCLE NUMBER and PUMP CURRENT pushbuttons together
for at least 2 seconds and the processor enters in a routine where it
is possible to program the controller.
In this routine, the CYCLE push-button is used for choosing/changing
the value or condition;the PUMP CURRENT push-button is used to
enter and confirm the value. At any time it is possible to exit this
routine by pressing the CYCLE and PUMP CURRENT pushbuttons at
the same time for at least 2 seconds.
The display shows:
C
S
and then
S
S
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to ramp-up
the Normal speed within ten steps. When NO is selected, the Soft
Start mode is deselected and the ramp-up of the pump will be done
within 300 seconds.
The controller factory default is set to YES.
278/310
O
N
F
I
G
O
F
T
W
A
X
X
X
X
O
F
T
S
E
L
E
C
T
Turbo-V 1000HT Controller User Manual / 87-900-894-01
U
R
A
T
I
R
E
V
E
X
X
X
X
X
T
A
R
T
I
O
N
:
O
N
R
S
I
O
N
X
X
M
O
D
E
X
X
X
(G)