Connection J6 and J7 – Accessories and
Options Interconnections
Figure 14
The 120 Vac, 1 A maximum (independent of line Voltage) output
Voltage is present when the main power switch is set to 1 position
and after START push-button is pressed and will remain present
until a fault condition is displayed on the front panel display or the
turbopump is stopped. The connector J6 is a power output for a
pump cooling fan, the connector J7 is for vent device, forepump
connector coils, etc. The maximum current that can be drawn from
J6+J7 is 1 A. To make connections, remove the plug and wire the pins
(maximum wire size 18 AWG, 1 mm²) as indicated in the figure to
obtain the desired capability.
On the J6+J7 outlets the output Voltage is 120 Vac with the maximum current of
CAUTION!
1 A. If more than one device is connected, be sure to draw not more than 120 VA
to avoid controller damage. Use a relay coil with a maximum surge power of
65 VA.
Forepump relay coil is an independent user supplied item.
NOTE
Turbo-V 1000HT Controller User Manual / 87-900-894-01
J6 and J7 connectors
Technical Information
Interconnections
(G)
17