The following parameters can be changed:
Tab. 8
BYTES
1
2
3-6
7
8
9
10
When the Request is "I" (Status readings), the Answer will contain a
string of 2 characters with the following parameters:
Tab. 9
BYTES
LSB X
MSB X
2
Turbo-V 1000HT Controller User Manual / 87-900-894-01
X
X
X
1
X
X
X
MEANING
Speed adjust value (in KRPM)
Speed threshold
Run up time in seconds
(long coded in 4 bytes)
Deat time
(0 = NO 1 = YES)
Soft Start mode
(0 = NO I=YES)
Water cooling
(0 = NO I=YES)
CRC
MEANING
0000 = STOP
0001 = WAITING INTERLOCK
0010 = STARTING
0011 = NORMAL OPERATION
0100 = HIGH LOAD
0101 = " "
" "
0110 = FAILURE
0111 = APPROACHING LOW SPEED
R2 status (0 = OFF 1 = ON)
R1 status (0 = OFF 1 = ON)
...
...
CRC