Operation
6
Operation
26 / 84
DANGER
A danger is posed by penetrating liquids.
There is an electric shock hazard for personnel.
If liquid has penetrated the incubator during oper-
ation, do not touch the incubator and switch it off
on the building side (e.g. via the fuse).
In such a case, contact a service technician of the
manufacturer.
WARNING
There is a risk of incorrect temperature control of the
samples.
There is a risk of incorrect temperature control of
microbiological samples, which can lead to misinter-
pretation of the test results or destruction of the sam-
ples.
Arrange for maintenance and calibration to
be carried out ⮫ 
on page 69 .
Note the information in this manual concerning
the effect of heat generation from auxiliary equip-
ment operated in the incubator.
Bear in mind the temperature change profile.
Do not place microbiological samples on the incu-
bator floor. Use the shelves.
Position the microbiological sample material within
the usable space.
Comply with the permissible ambient conditions.
Prevent the incubator from being exposed to
direct sunlight.
Close the incubator door when not inserting or
removing samples.
After a power failure, check to ensure that the
samples are still suitable for use.
CAUTION
A hazard is posed due to the improper pulling-out or
pushing-in of the various slide-in modules or drawers.
Avoid strong acceleration.
Carefully pull out or push in slide-in modules,
drawers or applications.
09.2023 / Rev. --
Chapter 7.5 'Maintenance'
AB69601en