5.2.2
Incubator connection
Power supply connection
AB69601en
1
Socket screw
2
Hole for fastening element
2.
By loosening the socket screw (1) on the fixing kit, adjust the
length at the wall distance.
3.
Retighten the socket screw (1).
WARNING
Risk of injury due to falling loads
Risk of injury due to insufficient pull-out resistance of
the fastening elements.
−
Select the fastening elements with a minimum
pull-out resistance of 200 N.
4.
Attach the fixing kit (2) to the wall using a suitable fastening
element (e.g. rawl plug and screw).
Alternatively another fastening option can be
selected.
However, ensure that the fastening is imple-
mented using 2 points at the top of the incu-
bator and that the fixation elements each have
a strength or pull-out resistance of 200 N.
DANGER
A hazard is posed by a damaged mains supply line.
There is an electric shock hazard for personnel.
−
Check the incubator and the mains supply line for
damage. If damage is discovered, the incubator
must not be operated.
−
The incubator may only be operated at a socket
with a personal protection switch.
09.2023 / Rev. --
Commissioning
1
2
21 / 84