Device description
3
Device description
3.1
Standard device
7
6
5
4
Fig. 1: Unit front side
1
Housing
2
Mains power switch
3
Device base
4
Door handle
5
Door lock
6
Touchscreen
7
USB port
14 / 84
Touchscreen including
■
Temperature protection
■
Tolerance range asymmetrically adjustable.
1
■
Door alarm individually adjustable.
2
The incubator is equipped with a temperature monitor of protection
class 3.1 in accordance with DIN12880:2007-05. The temperature
monitor is used to protect the incubator (device protection), its sur-
roundings and the sample material (sample protection) against imper-
missible temperature overshoots.
3
6
5
4
Fig. 2: Front of unit, door open
1
Temperature sensor
2
Inserting shelf
3
Interior made of high-quality stainless steel
4
Glass door (optional)
5
Glass door handle (optional)
6
Access port
For cable glands, use the access port on the side of the unit.
The incubator is fitted with a Ø 30-mm access port on the side of the
unit.
Cables from external measuring systems can be fed into the interior
through this opening.
After routing the cables through the opening, the
access port must be sealed on the unit's side to
prevent temperature deviations in the interior. If the
access port is not used, it must be sealed with a
plug.
If external devices are used in incubators without
cooling, overheating can quickly occur for which
the incubator cannot compensate. The over-temper-
ature can result in sample material being negatively
affected. Long-term over-temperatures can damage
the incubator.
When operating a thermal load inside the inner
space, monitor the device
09.2023 / Rev. --
1
2
3
AB69601en