±0.3°C for the rest.
Response time: ≤5s
Patient Group: Adult and Neonate
Measuring site: earhole
Deviation: ≤0.1°C
G. Operating Environment
Operating Temperature: 5°C ~40°C
Operating Humidity:
Atmospheric pressure:
Note: portable and mobile RF communications equipment may affect the perfor-
mance of the Oximeter.
H. Low Perfusion Performance
The accuracy of SpO
above when the modulation amplitude is as low as 0.4%.
I. Resistance to interference of surrounding light:
The difference between the SpO
light and that of darkroom is less than ±1%.
J. Wireless (bluetooth) function
Frequency band: 2.4GHz
Working profile: BLE V4.0
K. Dimensions: 158 mm (L) × 73 mm (W) × 25 mm (H)
Net Weight: about 230g (including battery)
L. Classification
Type of protection against electric shock:
Internally powered equipment and Class II.
Degree of protection:
Type BF applied parts.
Degree of protection against harmful ingress of liquids: The equipment is IP22
with protection against harmful solid foreign objects and ingress of liquid.
Mode of operation: Continuous operation.
Electro-Magnetic Compatibility: Group I, Class B
M. Data update period
The update time for determining SpO
ing update time is 1 second.
Remark: The oximeter calculates the SpO
recently acquired data segment, then yields the displaying value by moving aver-
age of the latest calculated parameters. The reading value of SpO
oximeter is updated every second, and the displayed plethysmogram is a normal-
ized waveform. If the signal is no integral (such as with too much noise, or poor
signal to noise ratio or signal is lost), then the SpO
invalid value, that's to say, the numeric reading will disappear and be displayed as
"--" instead.
Note: The oximeter is calibrated in the factory before sale, and there is no need for
user to calibrate again.
ENGLISH
15%~93%
70kPa~106kPa
and PR measurement still meet the precision described
2
value measured in the condition of indoor natural
2
2
and PR value is 8 seconds, and the display-
and PR value, every second by use of
2
and PR will be identified as an
2
62
and PR on the
2