Télécharger Imprimer la page

Gima SP-20 Mode D'emploi page 46

Oxymètre de pouls oxy 110
Masquer les pouces Voir aussi pour SP-20:

Publicité

Les langues disponibles

Les langues disponibles

 The fingernail should be of normal length.
 The SpO
sensor cannot be immersed into water, liquid or cleanser.
2
 The SpO
sensor can be repeatedly used. Please clean and disinfect before reuse.
2
 Connector with the label "SpO
connector with the label "TEMP" can only be connected with the temperature
probe.
3.2 Temperature Measurement (optional)
The ear thermometer probe is a delicate transducer. To operate please follow these
steps and procedures. Failure to accurately operate may cause damage to the probe.
The infrared temperature probe is as shown in figure 3.2.
Please place the ear thermometer probe in a stable ambient temperature for 30 min-
utes before taking a measurement.
Figure 3.2 the infrared temperature probe
Mode of operation of the clinic thermometer: Adjusted mode
Note: The default mode is ear temperature measurement.
Operation procedure:
1. Connect the ear thermometer probe to the connector on the upper side of device
marked with "TEMP". Press the measuring key. When the LCD screen of the device
displays " ", this indicates that the probe is successfully connected.
2. When the temperature unit "°C" on the screen of the probe is blinking, the user can
begin to take the measurement.
3. Take off the probe cover and insert the tip of the ear thermometer probe into the
earhole, Press the measuring key to start the measurement. A short beep means the
measurement has finished and the result will be displayed on the screen.
4. Press the unit key to switch between °C and °F.
Note:
 If the ear thermometer probe detects a hardware failure, a red screen will appear
ENGLISH
" can only be connected with SpO
2
1. Display screen
2. Unit key (°C/°F)
3. Measuring key
4. Measuring tip of the ear
thermometer probe
5. Battery cover
6. Temperature Probe Cover
46
probe, and
2

Publicité

loading

Ce manuel est également adapté pour:

34341 34341Oxy 11034341