on the display screen and the probe will not enter into measuring mode. Press the
measuring key to restart the measurement.
 The ear thermometer probe will switch to stand by automatically if there is no oper-
ation for 1 minute. If a further measurement is needed, press the measuring key and
repeat step 2 and step 3.
 Normal range of ear temperature: 35.8 ~ 38.0 °C
 Each person has his/her own normal temperature value, and the normal tempera-
ture value also changes at different time within a day. Therefore, it's recommended to
report your doctor not only the temperature value, but also the measuring position,
if possible you may provide your own normal temperature range to your doctor for
reference.
Safety Instruction for Temperature Measurement
 This device meets requirements established in ASTM Standard (E1965-98).
 Do NOT use the ear thermometer probe when the subject temperature and am-
bient temperature are outside the operating ranges specified by the manufacturer.
Performance of the device may be adversely affected when one or more of the
following occur:
A. Operation outside of the manufacturer specified subject temperature range.
B. Operation outside of the manufacturer specified operating temperature and hu-
midity ranges.
C. Storage outside of the manufacturer specified ambient temperature and humidity
ranges.
D. Mechanical shock.
 Manufacturer defined soiled or damaged infrared optical components.
 Do NOT take a measurement when the patient is moving.
 Patients with tympanitis and otitis problems should NOT use this device
4 Operation
4.1 Power on/off the Oximeter
• Long pressing "
powered on. The oximeter will do self-test and then the software version and warn-
ing message "Professional attendance is required for continuous monitoring!" will be
shown on the screen, as shown in figure 4.1 (refer to your oximeter for actual version).
" Power/Left key for 1~2 seconds, then the oximeter will be