• Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
• The cuffs are sensitive and must be handled with care.
• Only pump up the cuff once fitted.
• The function of this device may be compromised when used
close to strong electromagnetic fields such as mobile phones or
radio installations and we recommend a distance of at least
1 m. In cases where you suspect this to be unavoidable, please
verify if the device is working properly before use.
• Do not use this device if you think it is damaged or notice
anything unusual.
• Never open this device.
• If the device is not going to be used for a prolonged period the
batteries should be removed.
• Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
You can machine wash the cuff cover at 30°C (do not iron!).
WARNING: Under no circumstances, however, may you
wash the inner bladder! Always remove the sensitive
bladder from the sleeve before washing and replace it care-
fully again afterwards.
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
BP A80
All manuals and user guides at all-guides.com
10. Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
• Batteries, cuff and parts that become worn with use are not
• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
Please contact your local Microlife-Service (see foreword).
11. Technical Specifications
Operating conditions: 10 - 40 °C / 50 - 104 °F
Storage conditions:
Weight:
Dimensions:
Measuring procedure: oscillometric, corresponding to Korotkoff
Measurement range: 20 - 280 mmHg – blood pressure
Cuff pressure display
range:
Resolution:
Static accuracy:
Pulse accuracy:
Voltage source:
Battery lifetime:
IP Class:
Reference to
standards:
Expected service life: Device: 5 years or 10000 measurements
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
included.
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
15 - 95 % relative maximum humidity
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
610 g (including batteries)
151 x 115 x 88 mm
method: Phase I systolic, Phase V diastolic
40 - 200 beats per minute – pulse
0 - 299 mmHg
1 mmHg
pressure within ± 3 mmHg
± 5 % of the readout value
4 x 1.5 V alkaline batteries; size AA
approx. 3000 measurements
(using new batteries)
IP20
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Accessories: 2 years
EN
5