Télécharger Imprimer la page

arjo Alpha Active 4 Notice D'utilisation page 28

Masquer les pouces Voir aussi pour Alpha Active 4:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
Guidance and manufacturer's declaration - electromagnetic immunity
Immunity Test
IEC 60601-1-2
test level
Electrostatic
±2kV, ±4kV, ±8kV,
discharge (ESD)
±15kV air
EN 61000-4-2
±8kV contact
Conducted
3V in 0,15 MHz to
disturbances
80 MHz
inducted by RF
fields
6V in ISM and
amateur radio
EN 61000-4-6
bands between
0,15 MHz and
80 MHz
80% AM at 1 kHz
Radiated RF
Home Healthcare
environment 10 V/
electromagnetic
m
field
80 MHz to
2,7 GHz
EN 61000-4-3
80% AM at 1 kHz
Electrical fast
±1kV SIP/SOP
transient/burst
ports
EN 61000-4-4
±2kV AC port
100kHz repetition
frequency
Power frequency
30A/m
Magnetic field
EN 61000-4-8
Surge
±0,5kV ±1kV;
±2kV, AC Mains,
IEC 61000-4-5
Line to Ground
±0,5kV ±1kV, AC
Mains, Line to Line
Compliance
Electromagnetic environment -
Level
guidance
±2kV, ±4kV, ±8kV,
Floors should be wood, concrete or
±15kV air
ceramic tile. If floors are covered with
synthetic material the relative
±8kV contact
humidity level should be at least 30%
3V in 0,15 MHz to
Portable and mobile RF
80 MHz
communications equipment should
be used no closer to any part of the
6V in ISM and
product, including cables, than 1.0 m,
amateur radio
if the transmitter's output power rating
bands between
exceeds 1W
0,15 MHz and
80 MHz
Field strengths from fixed RF
transmitters, as determined by an
80% AM at 1 kHz
electromagnetic site survey, should
be less than the compliance level in
Home Healthcare
each frequency range
environment 10 V/m
Interference may occur in the vicinity
80 MHz to 2,7 GHz
of equipment marked with this
symbol:
80% AM at 1 kHz
±1kV SIP/SOP
Mains power supply should be that of
ports
a typical commercial or hospital
environment.
±2kV AC port
100kHz repetition
frequency
30A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
±0,5kV ±1kV;
Mains power quality should be that of
±2kV, AC Mains,
a typical commercial or hospital
Line to Ground
environment.
±0,5kV ±1kV, AC
Mains, Line to Line
22
(a)
(b)

Publicité

loading