2. Set treatment parameters
Before carrying out a treatment, set the parameters
based on the patient and the pathology. Select the
mode, the signal intensity level and the duration.
When an accessory with an electrode is connected,
the default mode corresponding to that electrode is
automatically activated, and only the modes that it
supports can be selected:
• When a mode is available, an illuminated line
appears under the mode.
• When a mode is not available, this line is not
illuminated and the entire column is grayed out.
• When a mode is available but not activated, it is
not highlighted.
• When a mode is activated, the entire column is
highlighted.
In the image below, only CET, RET, Hi-TENS and Hi-EMS modes can be selected (illuminated line), while CET, RET
and Hi-TENS modes are activated (columns are highlighted). MIX and Hi-TENS modes cannot be activated (modes
are grayed out).
With the TECAR X accessory, you can navigate on the interface.
The TECAR X accessory logo appears on the selectable mode:
• To change mode, swipe twice to the left or to the right.
• To activate/deactivate the mode, make a pronosupination movement with the wrist.
• To change the intensity of the treatment, tap the accessory with your finger. To increase the intensity by 10%,
tap 3 times, and to decrease the intensity by 10%, tap twice.
40
To use this function, please activate the "TIC" in the "Settings" menu.