5-50 μl pipettes:
1. Put the spring hatch 25, the spring 16, spring support 17 and the tube 18 back on the piston.
2. Slide the bigger O-ring 19, smaller O-ring 20 and O-ring support 21 on the piston.
3. Slide the small spring 22 on the piston.
4. Carefully slide the entire assembly into the tip, then push the spring hatch 25 down and
turn it 90 degrees.
5. Take the upper ejector top 11 between fi ngers and pull it down.
6. Push the tip cone into the handle while holding the ejector top 11 own, until you hear a "click".
7. Reassemble the lower tip ejector sleeve 13 (snap fi tting).
8. Press trigger to reconnect the piston to drive mechanism.
9. Press READY to return from service mode.
10-100 μl pipette:
1. Put the spring hatch 25, the spring 16, spring support 17 and the O-ring 20 back on the piston.
2. Carefully slide the entire assembly into the tip, then push the spring hatch 25 down and
turn it 90 degrees.
3. Take the upper ejector top 11 between fi ngers and pull it down.
4. Push the tip cone into the handle while holding the ejector top 11 down, until you hear a "click".
5. Reassemble the lower tip ejector sleeve 13 (snap fi tting).
6. Press trigger to reconnect the piston to drive mechanism.
7. Press READY to return from service mode.
30-300 μl pipette:
1. Put the spring hatch 25, the spring 16, spring support 17 and the O-ring 20 back on the piston.
2. Carefully slide the entire assembly into the tip, then push the spring hatch 25 down and
turn it 90 degrees.
3. Take the upper ejector top 11 between fi ngers and pull it down.
4. Push the tip cone into the handle while holding the ejector top 11 down, until you hear a "click".
5. Reassemble the lower tip ejector sleeve 13 (snap fi tting).
6. Press trigger to reconnect the piston to drive mechanism.
7. Press READY to return from service mode.
100-1000 μl pipettes:
1. Put the spring hatch 25, the spring 16, spring support 17 and the O-ring 20 back on the piston.
2. Carefully slide the entire assembly into the tip cone.
3. Push the spring hatch 25 down and turn it 90 degrees.
4. Take the upper ejector top 11 between fi ngers and pull it down.
5. Push the tip cone into the handle while holding the ejector top 11 down, until you hear a "click".
6. Reassemble the lower tip ejector sleeve 13 (snap fi tting).
7. Press trigger to reconnect the piston to drive mechanism.
8. Press READY to return from service mode.

Replacing the battery

1
2
There are two versions of battery lids in Finnpipette Novus Pipettes
- Follow the instructions below according to the battery lid version
Version 1
Opening the Version 1 battery lid
- Use a small fl at screwdriver for opening the battery lid.
- Put the head of the screwdriver into the slot and rotate it to open the lid.
- Disconnect the battery connector and pull out the battery.
Opening the Version 2 battery lid
- Open the battery lid by pressing and pulling the lid as shown in the picture.
- Disconnect the battery connector and pull out the battery.
See page 5 for installing the battery.
Note: Do not reuse version 1 style battery lid
14
Version 2