Installation

Section 3
The immersion circulator is designed for continuous operation and for
indoor use.
Never place the cooler in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Leave coolers in an upright position at room temperature (~25°C) for
24 hours before starting. This will ensure the lubrication oil has drained
back into the compressor.
The cooler requires clean air for proper operation. Air enters from the front
and exits through the rear.
The cooler construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
The cooler is intended for use on a dedicated outlet.
The cooler's power cord is the electrical disconnecting device, it must
be easily accessible at all times.
Operate the cooler using only the supplied line cords, never operate
equipment with damaged cords.
Refer to the nameplate on the rear of the cooler for specific electrical
requirements. Voltage range deviations of ± 10% are permissible. The outlet
must be rated as suitable for the total power consumption of the cooler.
Ensure the electrical cords do not come in contact with the evaporator
coil or hose.