Section 2 General Information
Specifications/
Equipment Ratings
Working Temperature
Ambient Temperature
Maximum Relative Humidity
(Non Condensing)
Operating Altitude
Pollution Degree
Overvoltage Category
Degree of Protection
Sound Level
Cooling Capacity at +20°C
at -10°C
at -40°C
at -60°C
Compressor Cooling
Refrigerant
Hose Length
Cooling Coil (Ø x L)
Smallest Bending Radius
Cooler Dimensions (H x W x D)
Evaporator Material
Nominal Weight kg/lb

Electrical Requirements

Total Maximum Wattage
2-2
Range
1
Range
2
Sea Level to 2000 meters (6560 feet)
3
350 watts/60Hz (300 watts/50Hz)
250 watts/60Hz (220 watts/50Hz)
4
5
1. The lowest possible temperature in a bath and the shortest cooling down time are dependent on: the
volume of liquid to be cooled, the type of liquid and its viscosity at the desired working temperature,
the heat conductivity of the liquid, the bath vessel and covering insulation, the heat created by any EK
immersed in the bath used to recirculate or stir the liquid.
The lowest temperatures possible for the EK45 and EK90 specified above were reached in a Dewar
vessel with a volume of 2 liters. Methylcyclohexane was used as the bath liquid and the bath was not
stirred.
2. The cooling capacity is reduced at ambient temperatures of > 25°C. 30°C max ambient for setpoints
>20°C for EK90, or >0°C for EK45. Example: Setpoint 40°C, max ambient = 30°C.
3. 1 meter from EK.
4. Never bend the probe when it is cold.
5. Nominal values ±10%.
• Thermo Fisher Scientific reserves the right to change specifications without notice.
EK45
-45°C to +40°C
-49°F to +104°F
+10°C to +40°C
+50°F to +104°F
80%/31°C to 50%/40°C
2
II
IP20
less than 58 dBA
-
-
Air
R404A
150 cm
81 x195 mm
-
380 x 225 x 465 mm
300 Series Stainless Steel
27.2 / 60.0
230V/50Hz or
115V/60Hz
900VA
EK90
-90°C to +40°C
-130°F to +104°F
+10°C to +40°C
+50°F to +104°F
80%/31°C to 50%/40°C
Sea Level to 2000 meters (6560 feet)
2
II
IP20
less than 58 dBA
300 watts
280 watts
170 watts
100 watts
Air
R404A/R23
150 cm
13 x 900 mm
40mm
490 x 380 x 460 mm
300 Series Stainless Steel
62.7 / 138.2
230V/50Hz or
115V/60Hz
1300VA
Thermo Scientific

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ek90

Table des Matières