6.2
Function checks
The function test must be performed before each use and after each intra-
operative change of battery and attachment.
►
Make certain there is a battery in the battery compartment.
►
Ensure that the cap has completely clicked into place.
►
Check that the attachment is coupled properly. To test this, pull on the
attachment.
►
Check that the tool is coupled properly. To do this, pull on the tool.
►
For GB891R: check that the tool attachment has locked into place. To
do this, turn the tool attachment.
►
Ensure that the tool blades are not mechanically damaged.
►
Release product for use (ON position).
►
Run the product shortly at maximum speed in right and left rotation.
►
Make certain that the rotational direction is correct in each case.
►
Check the product for damage, abnormal operation noises, heavy vibra-
tion and excessive heat.
►
Do not use the product if it is damaged or defective.
►
Set aside the product if it is damaged.
6.3
Safe operation
WARNING
Coagulation of patient tissue or risk of burns for patients and user
through hot product!
►
Cool the tool during operation.
►
Keep product/tool out of the reach of patients.
►
Let the product/tool cool down.
►
Use a cloth to protect against burns when changing the tool.
WARNING
Risk of infection from aerosol formation!
Risk of injuries caused by particles coming loose from the tool!
►
Use suitable protection (such as waterproof protective clothing,
face mask, safety gases, suction).
WARNING
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using the product.
WARNING
Risk of injury when using the product beyond the field of view!
►
Apply the product only under visual control.
WARNING
Risk of injury and damage to the tool/system!
The tool may get caught in drapes (such as textiles).
►
Do not let the tool come into contact with drapes (such as textiles)
during operation.
Note
The product drive motor is operated using a magnetic sensor system. In
order to prevent inadvertent activation of the motor, the product may not
be exposed to any magnetic fields (such as magnetic instrument pads).
Note
The slight whistling noise when the product is starting up is due to the con-
struction.
6.3.1
Normal Operation
Product operating in right rotation mode:
►
Activate the knob for speed control 2.
The speed of the product 1 is continuously regulated pursuant to the
attachment utilized.
Product operating in left rotation mode:
►
Press the knob for left rotation 11 completely in and keep it depressed.
►
Activate the knob for speed control 2.
The speed of the product 1 is continuously regulated pursuant to the
attachment utilized.
6.3.2
Tapping Operation or Oscillating Operation
Activate tapping operation or oscillating operation:
►
Keep the knob for left rotation 11 pressed for 3 seconds.
There will be multiple signal tones.
Tapping operation:
►
Depress the speed control knob 2 up to a maximum of half-way.
Product 1 will gradually turn counterclockwise and clockwise, whereby
the rotational angle in the clockwise direction is greater than that of
the counterclockwise direction.
Oscillating operation:
►
Press the knob for speed control 2 in all the way.
Product 1 will rotate clockwise and counterclockwise in alternation.
To activate left rotation:
►
Press the speed control knob 2 in addition to the knob for left
rotation 11.
The product 1 will rotate counterclockwise.
Deactivate tapping operation or oscillating operation:
►
Keep the knob for left rotation 11 pressed for 3 seconds.
There will be multiple signal tones.
en
7