Fault
Cause
Kirschner wire protection
Kirschner wire protection
sleeve cannot be
sleeve not compatible
attached
Connection on product
deformed/broken
Thread on Kirschner wire pro-
tection sleeve
deformed/defective
Rinsing adapter cannot
Rinsing adapter not compati-
be installed
ble
Connection on product
deformed/broken
Thread on rinsing adapter
deformed/broken
Cap cannot be
Lid not compatible
attached/removed
Cap deformed/defective
Connection to product
deformed/defective
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/broken
Battery shaft to product
deformed/defective
Sterile funnel or battery
Sterile funnel or battery
removal device not
removal device not compati-
attachable
ble
Sterile funnel or battery
removal device deformed/bro-
ken
Battery compartment on
product deformed/broken
Oil spray adapter not
Oil spray adapter not compat-
attachable
ible
Speed control knob not
Product in secured OFF posi-
working
tion
Speed control knob sticks/is
defective
Knob for counter-clock-
Left rotation knob sticks/is
wise rotation cannot be
defective
actuated
Recognition
Kirschner wire protection sleeve can-
not be screwed on
Kirschner wire protection sleeve can-
not be screwed on
Kirschner wire protection sleeve can-
not be screwed on
Rinsing adapter cannot be screwed on
Rinsing adapter cannot be screwed on
Rinsing adapter cannot be screwed on
Cap does not click into place
Lid is hard or impossible to
install/remove
Lid is hard or impossible to
install/remove
Battery cannot be completely inserted
into the battery shaft
Battery is difficult or cannot be
mounted/removed
Battery is hard or impossible to
insert/remove
Sterile funnel or battery removal device
not attachable to the battery shaft
Sterile funnel or battery removal device
difficult or not attachable
Sterile funnel or battery removal device
difficult or not attachable
Oil spray adapter not attachable
Safety catch is in OFF position
Speed control knob cannot be actuated Have the manufacturer repair the product.
Knob for counter-clockwise rotation
cannot be actuated
Remedy
Use the compatible Kirschner wire protection
sleeve for GA344.
Have the manufacturer repair the product.
Change Kirschner wire protection sleeve.
Use rinsing adapter compatible with GA344.
Have the manufacturer repair the product.
Replace the rinsing adapter.
Use the compatible cap for GA344.
Have the manufacturer repair the cap.
Have the manufacturer repair the product.
Use the compatible battery for GA344.
Have the manufacturer repair the battery.
Have the manufacturer repair the product.
Use compatible sterile funnel or battery removal
device for GA344.
Replace sterile funnel or battery removal device.
Have the manufacturer repair the product.
Use the compatible oil spray adapter for GA344.
Turn the safety catch to the ON position.
Have the manufacturer repair the product.