en
10. Technical Service
DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
►
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
►
Do not modify the product.
►
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
11. Accessories/Spare parts
Art. no.
Name
GA344244
Kirschner wire protection sleeve
GA344211
Rinsing adapter
GA675
Cover
GA346
Li-ion battery, short
GA678
Sterile funnel
GA679
Battery removal device
GB257R
ECCOS basket with holder
GB600
STERILIT Power Systems oil spray
GB600880
Oil spray adapter for GA344/GA844
TA011944
Cleaning brush
GB482R
ECCOS fixation for GA344
GB487R
ECCOS fixation lid
GB488R
ECCOS fixation sterile funnel
GB489R
ECCOS holder battery removal device
TA014550
Instructions for use of small drill GA344 (for ring binder)
TA014551
Instructions for use for small drill tool GA344 (flyer)
TA014552
Instructions for use for attachments for the small drill
tool GA344 and the drill tool GA844 (for ring-binder)
TA014553
Instructions for use for attachments for the small drill
tool GA344 and the drill tool GA844 (flyer)
14
12. Technical data
12.1 Classification according to Regulation (EU)
2017/745
Art. no.
Name
GA344
Small drill tool
12.2 Performance data, information about standards
Max. power
Max. motor speed
Direction of rotation
Cannulation
Weight (operational)
Dimensions (L x W x H,
ready for operation)
Application part
EMC
Conforming to standard
The product was subject to 500 preparation cycles as a test by the manu-
facturer and has passed.
12.3 Operating mode
Operation with non-periodic load and speed changes (type S9 pursuant to
IEC EN 60034-1)
Drilling (clockwise/counterclockwise rotation)
■
60 second application, 60 second pause
■
6 repetitions
■
30 min cooling time
■
Max. temperature 48 °C
Medullary reaming (clockwise/counterclockwise rotation):
■
30 second application, 30 second pause
■
8 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Drilling (oscillation)
■
15 second application, 15 second pause
■
3 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Saw mode with GB891R
■
30 second application, 60 second pause
■
4 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Saw mode with GB892R
■
30 second application, 60 second pause
■
5 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
approx. 85 W
1 250 rpm
Clockwise and counter-clockwise rotation,
oscillation
3.3 mm
0.84 kg ±10 %
177 mm x 124 mm x 55 mm ±5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
Class
IIa