en
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Fault
Cause
Product not running
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Product becomes too hot Excessive stress
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defec-
tive
Use of a defective attachment Product becomes hot
Blunt tool
Insufficient power
Product broken
Product is operating in left
rotation mode
Use of a defective attachment Attachment becomes very hot
Blunt tool
Loud running noise
Defective product drive/ball
bearings
Gearing/ball bearings of the
attachment broken
12
Recognition
No battery in the battery compartment Insert battery.
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product heating
Product becomes hot
Product heating
Tool and product become hot
Insufficient product performance
Large amount of heat generated after a
short time
Serrated tool runs anti-clockwise
Tool blades worn out
Loud, perceptible noise during opera-
tion
Loud, perceptible noise during opera-
tion
Remedy
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Observe instructions for use (nominal operating
mode).
Observe instructions for use (preparation, mainte-
nance).
Preventive measure: Oil product before every ster-
ilization.
Have the manufacturer repair the product.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Have the manufacturer repair the product
Operate serrated tool in a clockwise direction.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation, care).
Preventive measure: Oil product before every ster-
ilization.
Have the manufacturer repair the product
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every
sterilization.
Have the manufacturer repair the attachment