ALYA SWITcH LEVEr ASSEMBLY
- Insert lever "A" on switch pin.
- Place lever "A" hole in correspondance of "B" pin axe.
- Screw dowel "C " with the Allen wrench provided.
the control joystick must be handled delicately to
prevent breakage.
MAInTEnAncE
replacing the fuses "6" in the version with transformer (see the description of parts)
The light with transformer has two fuses of the same value.
They are replaced in the following way:
- Make sure that the power is off.
- Unscrew the plugs "6" on the transformer holder.
- Remove the fuses. After verifying the problem, replace them.
It is important that the new fuses have the specifications given on the data plate
and in the technical specifications.
ClEanInG tHE ParaBolaS "3" (see the description of parts)
For cleaning, Faro recommends using the detergent "Faro Perflex"; as an alternative, clean
with a soft cloth and isopropyl alcohol. Do not use detergents containing surfactants or
water repellent agents, whose build up can leave halos.
Hydroalcoholic disinfectants with 70% isopropyl or ethyl alcohol are suitable
for disinfection.
For cleaning plastic parts of the lamp ALYA, Do not uSE detergents-disinfectants
containing aMMonIuM HYDroXIDE - SoDIuM HYDroXIDE -
MEtHYlEnE CHlorIDE -MEtHYl alCoHol.
Faro has tested and suggests the following disinfectants: • Eco Jet-1 (Cattani
Group) • Sporekin Plus DS (Ims srl) • Zefirol Quick (Molteni Dental).
After disinfection, Faro recommends using the product "Faro Perflex" for
removing residue and halos.
PLEASE NOTE: other products might damage the reflectors (dishes). In case
of doubt, please contact CuStoMEr CarE Faro.
Slight halos do not impair the quality of light.
StErIlIzInG tHE HanDlES
To remove the handle, unscrew button "a" and slip it off. To insert
it, push firmly and screw on "a".
otHEr PartS oF tHE lIGHt (head-articulated arm)
Clean with a soft cloth.
It is strictly forbidden to use abrasive substances, detergents with trichloro-
ethylene, benzene, oil of turpentine or similar on any parts of the light.
Periodically perform the electrical safety check recommended for devices
installed in medical premises.
18
Handles are not provided sterile and must, therefore,
be sterilised before use.
Sterilize the device with standard cycles (SHPGRP) at
121°/134° C for a total of 200 sterilisation cycles.
B
A