The unit does not start or stop
The foot pedal does not activate the unit
The unit is still activated when the foot
pedal is released
is one acoustic signal and all LEDs are
blinking
The handpiece does not remain on the
magnetic support
The handpiece does not hold on the
handpiece cord
handpiece
98
Check the electrical connections and power supply at the mains
Check the fuses on back panel
Check that the foot pedal is properly connected
Replace the foot pedal
The switch in the pedal is blocked. Turn the unit off at the power switch
Replace the foot pedal
Powder chamber pressure is too low or too high
The air or liquid regulator is switched off
Install the magnetic support on a metallic part of the unit
Replace magnetic support
Check for presence of the O-ring seal and replace if necessary
Replace the handpiece
Replace the handpiece cord
Check the connections between the unit, the cord and the handpiece
Check for presence of water at the cord outlet port
Check the quantity of powder in your unit
Check for presence of powder/air at cord outlet port
Eliminate residual powder present in the channels using the cleaning needles provided