Mectron MULTIPIEZO PRO Mode D'emploi Et Maintenance page 47

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Problem
No liquid outflows
from the insert during
operation
The handpiece selection
lines are blinking
alternatively
Insufficient
performance
Possible cause
The insert type does not foresee
liquid passage
The insert is obstructed
The handpiece is obstructed
The irrigation level on the
screen is adjusted on "0"
The bottle is empty
The bottle is not correctly
installed
The silicon tubings of the pump
are incorrectly installed
The perisltatic pump is over-
used
One of the functions has been
activated and no handpiece has
been selected
The insert is not correctly
tightened on the handpiece
Insert broken, worn-out, or
deformed
12 Û TROUBLESHOOTING
Solution
Use an insert type with liquid
passage
Unscrew the insert from the
handpiece and free the insert
water passage by blowing
compressed through it.
If the problem persists, replace
the insert with a new one
Contact an Authorized Mectron
Service Center
Adjust the irrigation level
Fill in the bottle
Correctly connect the bottle to
the device body
Check the connections
of the silicon tubings
Replace the peristaltic pump
09.1)
(See paragraph
Select one handpiece by lifting it
from its support
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
(See paragraph 05.4)
Replace the insert with a new
one
EN