MULTIPIEZO PRO
Seal the wrench individually in a single-use
sterilization packet.
Sterilize the wrench in a steam autoclave.
EN
STERILIZATION METHOD - Torque wrench
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 10 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST 46:2002.
08.6 Û CLEANING OF THE BOTTLES
PREPARATION
- Disconnect the bottles from the device body
(see Chapter 07);
- Unscrew the caps from the bottles.
MATERIALS NECESSARY
- Water
- Detergent solution (pH 6-9). If need be, non-
aggressive disinfectant solution with neutral
pH (pH7)
- Clean and soft cloth with low fiber-release
- Sterile water
34
):
8
DANGER: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
WARNING: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
WARNING: do not exceed the allowed
load of the steam sterilizer.
WARNING: Do not sterilize the bottles
and their caps in autoclave, they could get
damaged.
WARNING: If you intend to disinfect,
we recommend that you use water-based
disinfectant solutions with a neutral pH
(pH7). Alcohol-based disinfectant solutions
and hydrogen peroxide are contra-
indicated, because they can fade the color
and/or damage the plastic materials. This
also holds true for chemical products such
as acetone and alcohol. Always rinse with
sterile water to preserve the disinfection.