On the keyboard, select the type of function
and irrigation necessary and the light, if
desired;
EN
At the end of the treatment position the
scaler handpiece on its support.
Only one handpiece can be used at a time,
it is therefore necessary to put the unused
handpice back to its place in order to enable
the other one to work.
NOTE: the multipiezo pro will recognize that
the handpiece is on its own support also by
placing only the cord properly.
05.5 Û IMPORTANT INFORMATION ON THE INSERTS
DANGER:
- When the titanium nitride coating is
visibly worn out, the insert must be
replaced. Use of an overly worn out insert
reduces its efficiency.
- Diamond-coated inserts: the diamond
inserts must be replaced when the
titanium nitride coating is visibly worn
out and, at any rate, after a maximum of
10 treatments.
- Do not activate the handpiece while
the insert is in contact with the treated
part, so that the electronic circuit can
recognize the best point of resonance
of the insert and allow its optimal
performance.
- Check the condition of wear of the insert
and that it is intact before and during
every use. In the event that a drop in
performance becomes apparent, proceed
to replace it.
- Use original Mectron inserts only. Use
of non-original inserts, in addition
20
9
10
to voiding the warranty, damages
the threading of the multipiezo pro
handpiece, with the risk of no longer
being able to screw the original inserts
correctly during subsequent use.
Moreover, the device settings are tested
and guaranteed to operate correctly
only when original Mectron inserts are
used.
- Do not change the shape of the insert in
any way by either bending or filing it. This
could cause it to break.
- Do not use an insert that has suffered
any type of deformation.
- Do not attempt to sharpen the insert
used.
- Always check that the threaded parts
of the insert and of the handpiece
are perfectly clean – see Chapter 08 -
CLEANING AND STERILIZATION.
- If excessive pressure is applied on the
insert, it can cause the insert to break
and possibly harm the patient.