08 Û CLEANING AND STERILIZATION
Rinse with tap water to eliminate all
3
residues;
Delicately brush the surface of the wrench
4
with a brush with soft nylon bristles using
distilled water;
Dry the wrench with a soft cloth
5
with low fiber-release;
Lubricate with medical degree lubricant in
6
the indicated point;
WARNING:do not use lubricants
based on oil or silicon.
Once you've finished the cleaning
7
operations, check the wrench under an
adequate source of light, paying attention
to the details that could conceal dirt
residues (cavities, grooves) and, if need be,
repeat the cleaning cycle;
EN
33