STERILIZATION METHOD - Inserts
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
):
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 20 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST 46:2002.
08 Û CLEANING AND STERILIZATION
7
WARNING: Before starting the
sterilization cycle, make sure that the
insert is thoroughly dry, both externally
and internally. For this purpose, blow
compressed air both externally and
through the internal thru hole; this will
prevent the onset of stains and haloes
on the surface or rust inside the insert.
Individually seal the inserts in single-use
8
sterilization packets.
Sterilize the inserts in a steam autoclave.
WARNING: Do not sterilize the insert
screwed onto the handpiece.
DANGER: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
WARNING: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
WARNING: Do not exceed the allowed
load of the steam sterilizer.
EN
31